ENTERING A TWO POINT CALIBRATION

The SMART3 Colorimeter can scan two reacted standards and create a

calibration curve. To prepare a calibration curve with multiple data points see

Entering a Multiple Calibration Curve (pg. 51).

1. From

the

User Test menu,

press

or

to

scroll to

STD (Standard)

Calibration.

111 H2O

Name the Test
Select Vial/WL

STD Calibration

Enter Constants

12:00:00 001/500

2. Press

ENTER

to select

STD

Calibration. The screen will

display Low Standard for

about 1 second and then

display the Low Standard

screen. A reverse font (dark

background with a light

character) will appear to

indicate the character that

will be adjusted. The same

character will also appear in

the center of the display.

Low Standard

9 . –

0

1 2 3

0

0.000000

12:00:00 001/500

3. Press

or

to scroll

to the fi rst character of the

low concentration. In this

example, 1.00 ppm.

Low Standard

. – 0

1

1 2 3

0

.000000

12:00:00 001/500

48

SMART3 Colorimeter 11.10