LaMotte SMART3 Colorimeter User Manual

Page 52

Advertising
background image

8. Press

ENTER

to save the

character. Repeat the

procedure until the low

concentration value is

complete. After the fi nal

character is complete the

meter will save the low

concentration value. The

screen will display

High

Standard for about 1 minute

and the meter will display

the

High Standard screen. A

reverse font (dark background

with a light character)

will appear to indicate

the character that will be

adjusted. The same character

will appear in the center of the

display.

High Standard

9 . –

0

1 2 3

0

.0000

12:00:00 001/500

9. Use

or

and

ENTER

to select the characters for the

high concentration value. In

this example, 7.5 ppm.

High Standard

9 . –

0

1 2 3

7.50000

0

12:00:00 001/500

10. After the fi nal character is

entered the meter will save

the high concentration value.

The screen will display

instructions for completing the

calibration procedure.

Insert Blank

<Enter> continue

12:00:00 001/500

50

SMART3 Colorimeter 11.10

Editing/Set Up