Hexavalent chromium procedure, Total chromium with acid digestion procedure – LaMotte SMART3 Colorimeter User Manual

Page 135

Advertising
background image

HEXAVALENT CHROMIUM PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 021 Chromium) from

TESTING MENU.

4. Scroll to and select 021 Chromium from menu.
5. Rinse a clean tube (0290) with sample water. Fill to 10 mL line with sample

water.

6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Use 0.1 g spoon (0699) to add one level

measure of *Chromium Reagent Powder (V-6276). Cap and shake for one

minute. Wait 3 minutes.

8. During the waiting period, fold a piece of fi lter paper in half, then in half again to

form a cone. Push corners together to open end, and insert into funnel (0459).

9. At the end of 3 minute waiting period, fi lter sample into a clean tube (0290).

Cap and mix. Insert tube into chamber, close lid and select SCAN SAMPLE.

Record result.

10. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

TOTAL CHROMIUM WITH ACID DIGESTION

PROCEDURE

1. Fill graduated cylinder (0418) to 50 mL line with sample water. Transfer to

Erlenmeyer fl ask (0431).

2. Use the 1 mL pipet (0354) to add 5 mL (fi ve measures) of *Sulfuric Acid, 5N

(7681). Swirl to mix.

NOTE: Highly buffered waters may require pH adjustment. Adjust the pH of

highly buffered samples to 7.0 ±0.5. Continue procedure.

3. Place

fl ask on burner or hot plate. Bring solution to a gentle boil.

4. Fill pipet (0341) with Potassium Permanganate, 0.5% (7682). While gently

swirling fl ask, add Potassium Permanganate, 0.5% (7682), 2 drops at a time

to boiling solution, until solution turns a dark pink color which persists for 10

minutes. Continue boiling.

5. Add one drop of *Sodium Azide, 5% (7683) to boiling solution. Boil for

approximately 30 seconds. If pink color does not fade, add another drop of

*Sodium Azide, 5%. Continue adding *Sodium Azide, 5% one drop at a time

until pink color disappears.

SMART3 Test Procedures 11.10

CHROMIUM, Hexavalent, Trivalent, Total

Test P

rocedures

Advertising