PROCEDURE
1. Press and hold
until colorimeter turns on.
2. Press
ENTER
to select TESTING MENU.
3. Select
ALL TESTS (or another sequence containing 027 Color) from
TESTING MENU.
4. Scroll to and select 027 Color from menu.
5. Rinse a tube (0290) with color-free water (distilled or deionized water). Fill to 10
mL line with color-free water.
6. Insert the tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Empty tube.
8. Rinse tube with sample water. Fill to 10 mL line with water sample.
9. Insert tube with sample water, close lid and select SCAN SAMPLE. Record
result in color units.
10. Press
to turn the colorimeter off or press
EXIT
to exit to a previous
menu or make another menu selection.
COLOR