Procedure – LaMotte SMART3 Colorimeter User Manual

Page 152

Advertising
background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 027 Color) from

TESTING MENU.

4. Scroll to and select 027 Color from menu.
5. Rinse a tube (0290) with color-free water (distilled or deionized water). Fill to 10

mL line with color-free water.

6. Insert the tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Empty tube.
8. Rinse tube with sample water. Fill to 10 mL line with water sample.
9. Insert tube with sample water, close lid and select SCAN SAMPLE. Record

result in color units.

10. Press

to turn the colorimeter off or press

EXIT

to exit to a previous

menu or make another menu selection.

COLOR

SMART3 Test Procedures 11.10