LaMotte SMART3 Colorimeter User Manual
Page 90
Advertising
APPLICATION:
Corrosion and rust inhibitors in cooling water systems
RANGE:
0.0 – 30.0 ppm Benzotriazole
0.0 – 30.0 ppm Tolyltriazole
MDL:
0.5 ppm Benzotriazole
0.5 ppm Tolyltriazole
METHOD:
Benzotriazole and tolyltriazole are UV-photolyzed in a
buffered solution with a pH between 4 and 6. A yellow
color develops in proportion to the concentration of triazole
present.
SAMPLE
HANDLING &
PRESERVATION:
Samples should be analyzed as soon as possible after
collection.
INTERFERENCES:
Tolyltriazole with interfere in the benzotriazole test.
Benzotriazole will interfere in the tolyltriazole test. Strong
reducing or oxidizing agents will interfere.
BENZOTRIAZOLE/TOLYTRIAZOLE
SMART3 Test Procedures 11.10