1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 002 Aluminum).

4. Scroll to and select 002 Aluminum from menu.
5. Rinse a clean colorimeter tube (0290) with sample water. Fill to the 10 mL line

with sample.

6. Insert tube into colorimeter chamber and select SCAN BLANK.
7. Rinse a clean test tube (0230) with sample water. Fill to the 5 mL line with

sample.

8. Remove tube from colorimeter. Empty sample from tube (0290).
9. Add 5 mL sample from test tube (0230) to empty tube (0290).
10. Use the 0.05 g spoon (0696) to add one measure of *Aluminum Inhibitor

Reagent (7865). Cap and mix to dissolve powder.

11. Use a 1.0 mL pipet (0354) to add 2 mL of *Aluminum Buffer Reagent (7866).

Cap and mix.

12. Use a second 1.0 mL pipet (0354) to add 1 mL of Aluminum Indicator Reagent

(7867). Cap and mix contents. Wait 5 minutes for maximum color development.

13. At end of 5 minute waiting period, mix, insert tube into chamber, close lid and

select SCAN SAMPLE. Record result.

14. Press

to turn the colorimeter off or press

EXIT

to exit to a previous

menu or make another menu selection.

NOTE: For the best possible results, a reagent blank should be determined to

account for any contribution to the test result by the reagent system. To determine

the reagent blank, follow the above test procedure to scan a distilled or deionized

water blank. Add 5 drops of Aluminum Complexing Reagent (7868). Then follow the

above procedure to perform the test on a distilled or deionized water sample. This

test result is the reagent blank. Subtract the reagent blank from all subsequent test

results of unknown samples. It is necessary to determine the reagent blank only

when a new lot number of reagents is obtained.