LaMotte SMART3 Colorimeter User Manual

Page 33

Advertising
background image

8. The displayed result can now

be standardized. Press

or

to scroll to calibrate.

002 Aluminum

0.28

ppm

%T/Abs

Calibrate

12:00:00 001/500

9. Press

ENTER

to select

Calibrate. A reverse font

(dark background with light

characters) will appear to

indicate that the reading can

be adjusted.

002 Aluminum

0.28

ppm

%T/Abs

Calibrate

12:00:00 001/500

10. Press

or

to scroll

to the concentration of the

prepared standard, 0.30 in

this example.

NOTE: A maximum

adjustment of 10% is possible.

If an adjustment of over 10%

is attempted, Overrange will

be displayed.

002 Aluminum

0.30

ppm

%T/Abs

Calibrate

12:00:00 001/500

11. Press

ENTER

to select

Calibrate. Two menu choices

will be offered, set calibration

and factory setting.

002 Aluminum

0.30

ppm

Set Calibration

Factory Setting

12:00:00 001/500

SMART3 Colorimeter 11.10

31