LaMotte SMART3 Colorimeter User Manual

Page 290

Advertising
background image

APPLICATION:

Drinking and surface waters, domestic and industrial

waste water.

RANGE:

0.00–3.00 ppm Zinc

MDL:

0.05 ppm

METHOD:

Zinc forms a blue colored complex with Zincon in a

solution buffered at pH 9.0. Other heavy metals are

complexed by cyanide and the zinc cyanide complex

is released by the addition of formaldehyde before the

other metal cyanide complexes are destroyed. Sodium

ascorbate is added to reduce the interference of

manganese.

SAMPLE HANDLING

& PRESERVATION:

Sample should be analyzed within 6 hours after

collection. The addition of hydrochloric acid will help

preserve the metal ion content, however the acid should

be neutralized before analysis.

INTERFERENCES:

The following ions interfere in concentrations greater than

those listed.

Ion

mg/L

Ion

mg/L

Cd(II)

1

Cr(III)

10

Al (III)

5

Ni(II)

20

Mn (II)

5

Co (II)

30

Fe (III)

7

CrO4(II)

50

Fe (II)

9

ZINC, Low Range

SMART3 Test Procedures 11.11

Test P

rocedures

Advertising