PROCEDURE - FRESH WATER

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 003 Ammonia-N LRF)

from TESTING MENU.

4. Scroll to and select 003 Ammonia-N LRF from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK. (See Note.)
7. Remove tube from colorimeter. Use the 1.0 mL plastic pipet (0354) to add 2.0

mL of *Salicylate Ammonia #1 (3978). Cap and mix.

8. Use the 0.15 g spoon (0727) to add two measures of *Salicylate #2 Reagent

(7457). Cap and mix until dissolved. Wait 1 minute.

9. At end of 1 minute waiting period use 0.1 g spoon (0699) to add two measures

of *Salicylate #3 Reagent Powder (7458). Cap and shake vigorously for at

least 30 seconds and all solid has dissolved. Wait 12 minutes for maximum

color development.

10. At the end of the 12 minute waiting period, immediately mix and insert tube into

chamber, close lid and select SCAN SAMPLE. Record result.

11. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

CALCULATIONS:

To express results as Unionized Ammonia (NH

3

):

ppm Unionized Ammonia (NH

3

) =

ppm Ammonia-Nitrogen (NH

3

–N) x 1.2

To express results as Ionized Ammonia (NH

4

):

ppm Ionized Ammonia (NH

4

+

) =

ppm Ammonia-Nitrogen (NH

3

–N) x 1.3

To determine the percentages of Unionized and Ionized Ammonia-Nitrogen, consult

the Appendix.

NOTE: It is strongly suggested that a reagent blank be determined to account for

any contribution to the test result by the reagent system. To determine the reagent

blank, follow the above test procedure to scan a distilled or deionized water blank.

Then follow the above procedure to perform the test on a distilled or deionized

water sample. This test result is the reagent blank. Subtract the reagent blank from

all subsequent test results of unknown samples. It is necessary to determine the

reagent blank only when a new lot number of reagents is obtained.

PROCEDURE - SALT WATER

AMMONIA NITROGEN, Low Range

SMART3 Test Procedures 11.10