LaMotte SMART3 Colorimeter User Manual

Page 276

Advertising
background image

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 086 Silica HR) from

TESTING MENU.

4. Scroll to and select 086 Silica HR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Add 6 drops *Silica Reagent #1 (V-4466). Cap

and invert to mix.

8. Add 12 drops of *Silica Reagent #2 (V-4467). Cap and mix. Wait 5 minutes.
9. At end of 5 minute waiting period, add 8 drops of *Silica Reagent #3 (V-4468).

Cap and mix.

10. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result.
11. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

NOTE: To extend the range to 100 ppm, perform a 2:1 dilution of water sample,

with silica-free water. Perform test and multiply result by 2.

SILICA, High Range

SMART3 Test Procedures 11.10

Test P

rocedures

Advertising