6. Insert the blank into the

chamber. Close the lid.Press

ENTER

to scan the blank. The

screen wil display

Blank

Done for about 1 second and

then return to the

Test Menu.

002 Aluminum

Scan Blank

Scan Sample

12:00:00 001/500

7. Insert the reacted sample

into the chamber. Close

the lid. Press

ENTER

to scan

the sample. The screen will

display

READING for about 1

second. The result will appear

on the screen.

002 Aluminum

1.00

ppm

Scan Blank

Scan Sample

12:00:00 001/500

8. To repeat the test, press

ENTER

to scan the sample again. The

last blank scaned is used by

the colorimeter for repeated

scans. A different blank can

be used by pressing

or

to scroll to Scan Blank

and then scanning another

blank. Scroll with

or

and make another

selection with

ENTER

. The %T

or Absorbance of the last test

can be viewed by choosing

%T/Abs. Press

EXIT

to

escape to previous menus.

NOTE: The menus loop in

this screen so either

or

will lead to the menu

selection needed.

002 Aluminum

1.00

ppm

Scan Bank

Scan Sample

12:00:00 001/500

28

SMART3 Colorimeter 11.10

Testing