Procedure – LaMotte SMART3 Colorimeter User Manual

Page 202

Advertising
background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 054 Lead) from TESTING

MENU.

4. Scroll to and select 054 Lead from menu.
5. Rinse a tube (0290) with sample water. Fill to 10 mL with sample.
6. Insert the tube into chamber, close lid and select SCAN BLANK.
7. Remove the tube from colorimeter. Empty the tube. Use the Syringe (0807) to

add 5mL of sample to the tube.

8. Add 5 mL *Ammonium Chloride Buffer (4032) to fi ll the tube to the 10 mL line.

Swirl to mix.

9. Add 3 drops *Sodium Cyanide, 10% (6565). Swirl to mix.
10. Use the 0.5 mL pipet (0353) to add 0.5 mL *PAR Indicator (4033). Swirl to mix.
11. Use the 0.5 mL pipet (0353) to add 0.5 mL Stabilizing Reagent (4022). Cap

and mix.

12. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result in

ppm as Reading A.

13. Remove tube from colorimeter. Add 3 drops *DDC Reagent (4034). Cap and

mix.

14. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result in

ppm as Reading B.

15. Calculate result:

Lead (ppm) = Reading A - Reading B

16. Press

to turn the colorimeter off or press

EXIT

to exit to a previous

menu or make another menu selection.

NOTE: For best possible results, a reagent blank should be determined to account

for any contribution to the test result by the reagent system. To determine the

reagent blank, follow the above test procedure to scan a distilled or deionized

water blank. Then follow the above procedure to perform the test on a distilled or

deionized water sample. This test result is the reagent blank. Subtract the reagent

blank from all subsequent test results of unknown samples. It is necessary to

determine the reagent blank only when a new lot number of reagents is obtained.

SMART3 Test Procedures 11.10

LEAD

Test P

rocedures

Advertising