Part names and functions, Cc-link communication unit lk-cc100, Part names and functions -4 – KEYENCE LK-CC100/DN100 User Manual

Page 10: Cc-link communication unit, Lk-cc100 -4, L run (operating status) indicator, Sd (send data) indicator, Rd (receive data) indicator, L err (error) indicator, Station no. setting switch

Advertising
background image

1

1-4

1 Before Use

Part Names and Functions

This section describes the name and functions of each component.

CC-Link communication unit LK-CC100

L RUN (Operating status) indicator

Lit: Data communication between the master and local

stations (when lit green).

Unlit: Data communications timed out (indicator will

light up again when data is received normally).

SD (Send data) indicator

Lit: Sending data.

RD (Receive data) indicator

Lit: Receiving data.

L ERR (error) indicator

Lit: A communications error exists (when lit red).

Flashing at regular intervals: Station number or

communication speed change detected while power

was on.

Flashing irregularly: Terminator (terminating resistor) is set incorrectly, or noise is

affecting the unit or CC-Link dedicated cable.

Unlit: Communications are normal.

STATION No. setting switch

Sets the station number (setting range: 1 to 64). x10 indicates the tens place, x1 indicates

the ones place.

B RATE setting switch

Sets the baud rate. The baud rate settings can be set as follows.

0: 156 kbps, 1: 625 kbps, 2: 2.5 Mbps, 3: 5 Mbps, 4: 10 Mbps

MODE setting switch

Sets the communication mode. The communication mode settings can be set as follows.

Switch setting

Multiple

Station

Version

Remark

0

1

1

-

Not supported (Disables control input, tolerance comparison
and readout of measurement values for any OUT).

1

1

2

1.1

Disables control input, tolerance comparison and readout of
measurement values for OUT03 and higher.

2

1

3

1.1

Disables control input, tolerance comparison and readout of
measurement values for OUT05 and higher.

L RUN

SD

RD

L ERR

10

LK-CC100

0

9 8 7

6

5

4

3

2

1

1

0

9 8 7

6

5

4

3

2

1

B RATE

SLD

DG

DB

DA

0

9 8 7

6

5

4

3

2

1

MODE

0 F

E D C B A

9 8

76

5

4

3

21

ST

A

TION No

.

(FG)




Advertising