Grounding the scale – A&D Weighing HW-1200KGV3 User Manual

Page 18

Advertising
background image

Page 16

HV-G/HW-G Series

5. Grounding the scale

5.4. Grounding the scale

When using where there may be static electricity, ground the scale.
The grounding procedure depends on the scale model. Refer to the table below.
These procedures are only for grounding part of the scale.

Models

Refer to

HV-15KGL/HV-15KGV/ HW-10KGL/HW-10KGV

Procedure A

HV-60KGL/HV-60KGV/HV-200KGL/HV-200KGV/HW-60KGL/HW-60KGV
HW-100KGL/HW-100KGV/HW-200KGL/HW-200KGV

Procedure B

HW-300KGL4/HW-300KGV4/ HW-600KGL4/HW-600KGV4

Procedure C

HW-600KGL3/HW-600KGV3/ HW-1200KGL3/HW-1200KGV3

Procedure D

Procedure A
(HV-15KGL/HV-15KGV/ HW-10KGL/HW-10KGV)

Secure the grounding cable using a M4 screw in the

screw hole between the two hexagon bolts on the
base unit bottom side. (Part of “ ”)



Procedure B

HV-60KGL/HV-60KGV/HV-200KGL/HV-200KGV/HW-60KGL/
HW-60KGV/HW-100KGL/HW-100KGV/HW-200KGL/HW-200KGV

Secure the grounding cable using the screw that

secures the under cover. (Part of “ ”)


Procedure C
(HW-300KGL4/HW-300KGV4/ HW-600KGL4/HW-600KGV4)

Remove the pan. Secure the grounding cable using
the screw that secures the band of the load cell cable
bundle. (Part of “ ”)
Pull the grounding cable through to the lower section
through the channel. Be careful that the pan does not
touch the grounding cable.



Procedure D

(HW-600KGL3/HW-600KGV3/HW-1200KGL3/HW-1200KGV3)

Remove the pan. Secure the grounding cable using
the screw that secures the summing box. (Part of “ ”)
Remove the cover located on the base’s side. Pull the
grounding cable out through this hole. Be careful that
the pan does not touch the grounding cable.


Load cell

cable

Summing box

Cover

Base unit

bottom side

Levering feet

Under cover

Base unit

bottom side

Pull the grounding

cable through to the

under side of the base.