Complete calibration procedure, Gravity acceleration correction, Preparation – A&D Weighing HW-1200KGV3 User Manual

Page 51

Advertising
background image

HV-G/HW-G Series

Page 49

14. Calibration (Adjusting Scale)

14.2. Complete Calibration Procedure

14.2.1. Gravity Acceleration Correction

Step 1 Turn on the display.

Open the rear cover of the display unit. Locate the

CAL switch inside.

Press and hold the

CAL

switch to enter the calibration

mode.

Then

Cal 0

is displayed.




Step 2 Press the

ZERO

key to enter the gravity acceleration

correction mode.



Step 3 Set your local gravity acceleration using the following

keys.

key

Selecting the figure to be changed.

key

Changing the number of the figure.



Step 4 Press the

ENTER

key to store the new value.

Then

Cal 0

is displayed.



Step 5 Press the

CAL

switch again. Then the scale returns to

the normal weighing mode.


14.2.2. Preparation

Step 6 Confirm the environmental conditions as follows:

Maintain a constant temperature and stable power.

Install the scale on a solid floor where there is no draft, vibration, strong magetic

fields or direct sunlight.

Refer to "4.1. Installing the Scale".


Step 7 Keep the display turned on for at least 30 minutes to warm up the scale.

Advertising