Command mode, Command list – A&D Weighing HW-1200KGV3 User Manual

Page 62

Advertising
background image

Page 60

HV-G/HW-G Series

16. RS-232C Serial Interface

16.3. Command Mode

The command mode is the function which can perform "outputting data", "controlling

the scale" and "setting parameters" by a command transmitted from a computer

Caution

Allow at least 500 milliseconds between commands.


16.3.1. Command List

The following explanation uses “Format 1 ( f15 0 )"

Data output

The current weighing data is output.

Template Q
Command

Reply


Selection of mode and unit

Selects the mode and unit. This is the same as the

MODE

key.

Template U
Command

Reply

The scale changes mode and unit.


Zero

The current mass value is set to the zero point.

This is the same as the

ZERO

key.

Template Z
Command

Reply

The mass value becomes zero and the zero point mark is displayed.


Tare

With a tare (container) placed, the current mass value is set to zero, and the net is
displayed. This is the same as the

TARE

key.

Template T
Command

Reply

The current mass value becomes zero and the net mark is displayed.


Cancel of tare

The weight value becomes the gross and the net mark is turned off. (The tare value
becomes zero.)

Template CT
Command

Reply

Gross is displayed and the net mark is turned off.