HV-G/HW-G Series

Page 59

16. RS-232C Serial Interface

16.2. Stream Mode

The scale outputs the current weighing data at the time the display is refreshed.

The scale does not output data while in the setting mode.

Averaging of function table

Refresh rate

A

f13 0 Fast response

Sensitive value

Approximately 7 times/s while the display is unstable,

Approximately 4 times/s when the display is stable

A

f13 1

Ú

A

f

f13 2 Slow response

Stable

value

Approximately 4 times/s

A


16.2.1. Preparation and Performance

(Examples)

Step 1 Turn off the display.

Press and hold the

ZERO

key and press the

ON/OFF

key. The function table is displayed.





Step 2 Press the

ENTER

key to display f4 ("Baud rate").

Select a parameter of "Baud rate" with the

key.




Step 3 Press the

ENTER

key to display f5 ("Output

mode").

Select "Stream mode ( f5 0 )" with the

key.


Step 4 Press the

ENTER

key to display f15 ("Format").

Select "Format 1 ( f15 0 )" with the

key.



Step 5 Press the

ENTER

key to display end .



Step 6 Press the

ENTER

key to store the new parameters.

The scale returns to the normal weighing mode and
outputs the data continuously.