Preparation (setting parameters), A a a – A&D Weighing HW-1200KGV3 User Manual

Page 36

Advertising
background image

Page 34

HV-G/HW-G Series

10. Accumulation Function

Function table

Description

A

f8 3

When the display is a positive stable value and not nearly-zero, the
scale accumulates the data automatically. The next accumulation
can be performed after the display becomes nearly-zero or a
negative value.

f8 4

When the display is a stable value and not nearly-zero, the scale
accumulates the data automatically. The next accumulation can be
performed after the display becomes nearly-zero.
Use

Recording the number and mass of articles removed from
the pan. (Place the articles on the pan. Press the

TARE

key

at each removal.)

A
A
A
A
A

f8 5

At each end of the full/dribble batch function, the scale accumulates
the data automatically.
Use

Packaging articles like a powder, it is used for recording the
number of bags and the total mass.

A
A
A


10.1. Preparation (Setting Parameters)

Step 1 Turn off the display.

Press and hold the

ZERO

key and press the

ON/OFF

key. The function table is displayed.




Step 2 Press the

ENTER

key to display an item of the

accumulation function (

f8

).




Step 3 Select a parameter of the accumulation function

(

f8 1

to

f8 4

) with the

key.



Step 4 Press the

ENTER

key and the

F

key to display

end . Press the

ENTER

key to return to the

normal weighing mode.