14. Calibration (Adjusting Scale)

14.2.3. Calibration of the Zero Point

Step 8 After 30-minute warm up, press and hold the

CAL

switch to enter the calibration mode.

Then

the

Cal 0

is displayed.


Step 9 Confirm that nothing is placed on the pan.

Wait for the stability mark to be displayed.
Press the

ENTER

key.

The scale stores the current condition as

the zero point.


Step 10 The scale displays

5pn1

for several

seconds. Calibration of the zero point is
finished. To exit the calibration mode,
proceed to step 14.

14.2.4. Span Calibration

Step 11 Set the value of the calibration mass

using the following keys. (The initial value
depends on the scale model.)

key

Selecting the figure to be

changed.

key

Changing the number of

the figure.


Step 12 Place the mass on the pan which was set

at step 11. Wait for the stability mark to be
displayed. Press the

ENTER

key.

The scale then calculates the span and

stores it.


Step 13 The scale displays

end

.

Remove the mass from the pan.


Step 14 Press the

CAL

switch to return to the normal

weighing mode.