HV-G/HW-G Series

11. Upper/Lower Comparator Function

11. Upper/Lower Comparator Function

This function compares a weight value with the upper limit (HI) and the lower limit

(LO) and displays the results.

To use this function, set the "Comparator function (

f6 0

to

f6 7

)" parameters in the

function table, upper limit value (HI) and lower limit value (LO).

Install option OP-03 or OP-04 to use relay output of the comparator.
Install option OP-03 to use buzzer output of the comparator.
The settings are stored in the scale even if the power is removed. (Refer to Caution

on the next page.)

Comparator Indicators

Comparison results are displayed by indicators

or


Comparison Condition

Weight

value < Lower limit value..................................LO is displayed and output.

Lower limit value ≦ Weight value ≦ Upper limit value......OK is displayed and output.

Upper limit value < Weight value ........................................HI is displayed and output.


Parameter List and Word Definition

The decimal point is not displayed in the upper or lower limit value settings.
The " nearly-zero " is within ±5 digits from the zero point in the unit of kg.


Function table

Description

A

f6 0

Press the

SET

key to compare the current weight value always.

Press the

SET

key again to stop the comparison.

A
A

f6 1

The scale always compares the weight value.

A

f6 2

Press the

SET

key to always compare the weight value when not

nearly-zero.
Press the

SET

key again to stop the comparison.

A
A

f6 3

The scale always compares the weight value when not nearly-zero.

A

f6 4

When the weight value becomes stable after the

SET

key is

pressed, the scale compares the weight value. It does not compare on
an unstable condition. Press the

SET

key again to stop the

comparison.

A
A
A
A

f6 5

When the weight value is stable, the scale compares the weight
value. It does not compare on an unstable condition.

A
A

f6 6

When the weight value becomes stable and not nearly-zero after
pressing the

SET

key, the scale compares the weight value. It does

not compare on an unstable condition.
Press the

SET

key again to stop the comparison.

A
A
A