Key operation

Display example

Description

STEPS

M2:

FIXED CURR

[email protected]

Move the cursor to the desired digit using the [SEL] key

and change the numeric value using the [UP] key.

Repeat this operation to change the value to 12.0mA.

[SEL]<^[UP1

mA

When the value has changed to the desired

[UP] [SEL] [SET]

numeric value, press the [SET] key to set the value

temporarily.

STEP?

M2:

FIXED CURR

12.0

When you press the [SET] key, the selected item is set

temporarily and a message confirming you whether the

setting is OK is displayed.

[SET]

/

OK?

[NO] ÍSET1

If OK, press the [SET] key. If you want to redo the

setting, press the [NO] key.

STEP8-1

[NO]

M2:

FIXED CURR

@0.0

mA

[UPl [SEL] [SET]

Pressing the [NO] key when you are asked "OK?"

causes the numeric value to return to the previous

value and enables you to redo the setting.

STEP8-2

M3:

FIXED PULSE

Soooo

Pressing the [SET] key when you are asked "OK?"

causes the fixed output current value setting to be

determined and the fixed output pulse setting screen to

[SET]

PPS

be developed automatically
At the same time, the cursor appears.

[EXT] [ENT]

(The digit on which the cursor is positioned is

reverse-displayed.)

STEP9

M3:

FIXED PULSE

ooQoo

Move the cursor to the desired digit using the [SEL] key

and change the numeric value using the [UP] key.

Repeat this operation to change the value to 100 PPS.

[SEL]^[UP]

PPS

When the value has changed to the desired numeric

[U P] [SELl [SETl

value, press the [SET] key to set the value temporarily.

STEP10

M3:

FIXED PULSE

00100

When you press the [SET] key, a message confirming

you whether the setting is OK is displayed.

If OK, press the [SET] key. If you want to redo the

[SET]

OK?

[NO] [SET]

setting, press the [NO] key.

STEP11-1

[NO]

M3:

FIXED PULSE

Soooo

PPS

[UP] [SEL] [SET]

Pressing the [No] key when you are asked "OK?"

causes the numeric value to return to the previous

value and enables you to redo the setting.

STEP11-2(=END)

M3:

FIXED PULSE

00100

Pressing the [SET] key when you are asked "OK?"

causes the setting value to be determined.

Then press the [EXT] key. You return to the menu

[SET]

PPS

[EXTl [ENTl

screen.

Note 1:

If you set a value beyond the allowable range, 2.4mA or 24mA (in the case of fixed

current output) or 24mA or lOOOOpps (in the case of fixed pulse output) is forcibly set.

Note 2; The pulse width set in Section 8.2.13 is used for fixed pulse output. The pulse width

must not be greater than one half of the fixed output setting frequency. However, the
pulse width will be 40% of fixed output setting frequency if it exceeds lOOOpps.

Note 3:

If the fixed output is set to ON, the fixed output current value and fixed output pulse

value setting screen is automatically developed. However, fixed output actually starts
when the fixed output pulse value setting is determined. (If the fixed output current
value and fixed output pulse value are set independently, fixed output starts when the
relevant setting is determined independently.)

-