The following example shows how to change the zero offset value from +1.0% to -2.5%.

Key operation

Display example

Description

N:

MEAS MODE

^MANUAL ZERO

Select "MANUAL ZERO" in the configuration item

selection screen.

fSELl fENTl

STEP1

N1:

MANUAL ZERO

-K)01.0

The current setting value (+1.0% in this example) is

displayed.

[ENT]

%

Then press the [ENT] key.

[EXT] [ENT]

* Pressing the [EXT] key returns you to the menu

screen.

STEP2

[ENT]

N1:

MANUAL ZERO

¡¡001.0

%

The switch name display at the bottom changes

(three keys {[UP], [SEL] and [SET]).

At the same time, the cursor appears.
(The digit on which the cursor is positioned is

reverse-displayed.)

[U P] [SEL] [SET]

STEPS

[SEL]0[UP]

N1:

MANUAL ZERO

-002.1

%

Move the cursor to the desired digit using the [SEL]
key and change the numeric value using the [UP]

key.
Repeat this operation to change the value to -2.5%.

[U P] [SEL] [SET]

When the value has changed to the desired

numeric value, press the [SET] key to set it

temporarily.

STEP4

[SET]

N1:

MANUAL ZERO

-002.5

%

OK? ®

When you press the [SET] key, a message
confirming you whether the setting is OK is

displayed.

If OK, press the [SET] key. If you want to redo the

setting, press the [NO] key.

[NO] [SET]

STEP5-1

[NO]

N1:

MANUAL ZERO

¡¡001.0

%

Pressing the [NO] key when you are asked "OK?"

causes the numeric value to return to the previous

value and enables you to redo the setting.

[U P] [SEL] [SET]

STEP5-2(=END)

[SET]

N1:

MANUAL ZERO

-002.5

%

[EXT] [ENT]

Pressing the [SET] key when you are asked "OK?"

causes the setting value to be determined.

Then press the [EXT] key. You return to the menu
screen.

Pressing the [ENT] key enables you to restart

setting value change operation from a state of
+2.5%.

Note; The manual zero adjustment value can be set within a range of ± 1 m/s equivalent (± 10 % of

the maximum range). If you set a value out of this range, an error message * HIGH OVER

SPEC * or * LOW OVER SPEC * is displayed. In this case, redo the setting.

In addition, you perform zero adjustment with one-push key while the detector is filled with

still water; the zero offset adjustment value will be automatically cleared to 0.0%.

-