Notes on wiring

A CAUTION

(

1

)

(

2

)

(

3

)

(4)

(5)

LF602F

Select the cable runs away from electrical equipment (motors, transformers, or radio
transmitters) which causes electromagnetic or electrostatic interference.

Deterioration of flowmeter circuit insulation occurs if the converter interior or cable ends get

wet or humidified. This in turn causes malfunction of flowmeter or noise problems. Avoid a

rainy day if the flowmeter is to be installed outdoors. Even indoors, prevent water from

splashing over the flowmeter. Try to finish the wiring as quickly as possible

The converter has an arrestor installed inside. Therefore, do not conduct a withstand voltage
test for the converter. To check the insulation of the converter, use a voltage of 250Vdc or

less.

After wiring, be sure to install the terminal block protection cover.

Because the excitation cable and flow rate signal cable transmit very delicate signals, pass each

of them separately through a thick steel conduit tube, keep them away from the large current

wiring as far as possible, and do not install them in parallel.

5.1 Cables

Use the kind of cables shown in Table 5.1 to wire the converter.

Table 5.1 Installation Cables

Name

Cable name

Nominal

cross-sectional

area

Finished

outer

diameter

Description

Power cable

3-core vinyl sheathed cable or
2-core vinyl sheathed cable

2 mm^

11~13mm

CW-JISC3401,

IEC60695,
IEC60754,
IEC60227,
IEC60245

or equivalent

Output signal

cable

The number of conductors the cable contains differs depending on
the specification of the output signal cable.

Use a shielded cable of finished outer diameter 11 to 13mm and
nominal cross-sectional area 1.25mm2.

CW-SJIS-258-C

or equivalent

Flow rate

signal cable

2-core shielded chloroprene
cabtyre cable

0.75 mm2

11~13m

m

2PNCT-S

JIS C 3327
or equivalent

Excitation

cable

3-core shielded chloroprene
cabtypre cable

2 mm2

1.25 m2

11~13m

m

2PNCT

JIS C 3327
or equivalent

LF602F

LF602F

-

22

-