10.9 Multiple range high/low limit alarm function (option)

The procedure to use multiple range high/low limit alarm is shown below.

Range setting

Set the range in accordance with 8.2.13

Span (range)

in the following order.

1. Set the range type to

“uunidirectional flow, multiple ranges selected

by external signal”.

2. Set Range 1 and Range 2 respectively.

DO setting (Note 1)

Set digital output 1 and 2 (DOl, D02) function to

multiple range high limit

alarm output and multiple range low limit alarm output

respectively in

accordance with

8.2.20 Digital I/O.

Also set the alarm output state to either of

normally open

or

normally close.

DI setting

Set digita

Digital I/O.

input (DI) function to RANGE SW in accordance with 8.2.20

High/Low limit

alarm value setting

Set

high limit alarm and low limit alarm

to

ON

and set alarm value to high

and low alarms respectively in accordancewith 8.2.24

Flow Rate High, Low,

High-High and Low-Low limit Alarm Setting.

Set the alarm not to use to OFF.

High High/Low Low limit

alarm value setting

• Set

high-high alarm and low-low alarm to ON

and set alarm value to them

respectively in accordance with

8.2.24 Flow Rate High, Low, High-High

and Low-Low limit Alarm Setting.

Set the alarm not to use to OFF.

(Note

1) When setting DO using HHT AF900 (Ver2.40 or older),

set alarm outputs

as below:

Multiple range high limit alarm output SPECIAL-B,
Multiple range low limit alarm output SPECIAL-A

-