Totalizer operation, Using control keys on the panel (option), Toshiba – Toshiba LF600F User Manual

Page 119

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

TOSHIBA

6,F,8A0i8|6|9,

Totalizer Operation

• Using control keys on the panel (option)

To start, stop or clear (reset) the totalizer, follow the procedure described below:

Key operation

Display example

Description

FLOW

2.000

m/s

RANGE1

5.000

m/s

fCNTl fSETl

The measured value is displayed (measurement

mode).

[CNT]

COUNT CTRL

TOTAL CNT

100

FOR
[STA] [CLR] [EXT]

Pressing the [CNT] key when you are in the

measurement mode causes the screen to switch to
the total counter operation screen.

In the sub display area of this screen, the total

counter value (both sides) is automatically

displayed.

In addition, the [STA], [CLR] and [EXT] keys are

displayed.

* When the total counter is operating, [STP] is

displayed. While the total counter is stopped, [STA] is displayed.

* If

you enter a wrong password at password

setting, the [CLR] key is not displayed.

[STA]

COUNT CTRL

TOTAL CNT

100

FOR CNT
[STPl [CLR] [EXT]

Pressing the [STA] key causes the total counting to

start and "CNT" to be displayed on the screen. In

addition, the [STA] key changes to the [STP] key.

[CLR]

COUNT CTRL

TOTAL CNT

0

FOR CNT
[STP] [CLR] [EXT]

Pressing the [CLR] key causes the total counter

value to be cleared.

[EXT]

FLOW

2.000

m/s

RANGE1

5.000

m/s

Pressing the [EXT] key causes you to return to the

measurement mode.

Notes: 1. Since the flow direction code is B (bidirectional flow (Forward/Reverse directions) automatic

switching),
• When you select forward/reverse multi-range,

the forward direction totalized value (count value) is displayed for operation in the
forward direction range.
the reverse direction totalized value (count value) is displayed for operation in reverse
direction range.

2. When you reset the counter, the flow counts for both directions will be cleared to zero.

3. Non-volatile memory is used to store the totalizer counter value. Therefore, the value will be

retained in the memory even if the power is cut off.

-

118

-

Advertising
This manual is related to the following products: