15. Specifications

15.1 Specifications

■ General Specifications

Measuring range: (measuring range by flow rate conversion)

0-0.3m/s to O-lOm/s

(A range of O-O.lm/s to 0-0.3m/s can be dealt with by an option specified at order time.)

Accuracy: (Accuracy when combined with the detector)

Flow rate to the

range (%)

20%

20

-

100

%

0 - 50%

50 -100%

Accuracy

0.1

less than 0.3 m/s 0.3 — less than 1.0 m/s

±0.25 %FS

±0.5% of rate

±0.25 %FS

±0.5% of rate

1.010 m/s

±0.1 %FS

±0.5% of rate

(Note) Accuracy under the basic operation conditions with Toshiba calibration facility)

Conductivity

: 5

¡

jl

S/cm or more

Ambient temperature:

-20

to +60°C

Storage temperature : -25 to +65°C

Power supply : 100 to 240Vac (allowable voltage range: 80 to 250Vac 50/60Hz)

24Vdc

(allowable voltage range: 18 to 36Vdc) or

1 lOVdc (allowable voltage range:90 to 130Vdc)

Power consumption : 27VA (17W) or less

Input

Input signal: • Flow rate proportional signal from the detector

• Digital input signal (option)

Signal type:

20 to 30Vdc voltage signal

Input resistance:

About 2.7k Q

Number of input points:

1

Digital input function (option): Select either of the following.

Range switching input: Large/Small range switching of unidirectional double range,
forward/reverse direction double range

Counter control input: Internal totalization counter start/stop/reset control

• Output hold input: The current output and pulse output are kept to their preset

values.

• Zero adjustment input: Start still water zero adjustment.

-

150

-