3. Example for pulse width:

The pulse width can be set from 0.3ms to 500ms in increments of 1ms. However,
the pulse width should be set to less than 40% of the pulse rate due to the setting
range and counting rate.

If “0” is set, the pulse width automatically will be set to 40% of the pulse rate

(100ms max.)

Example 1

Case Range

Counting rate(pulse rate)

Example2

Example3

:3600m^/h (ImVs)
lO.OOlm^

the pulse rate

: 3600(m^/h) / 0.00l(m^) =36000000 pulses/h

=1000pulses/p

the pulse rate for full scale = 1ms

*the pulse width = 1ms x 40% = 0.4ms

Case Range

Counting rate (pulse rate)

:3600m^/h
:1000m^

(ImVs)

the pulse rate

: 3600(m^/h) / lOOO(m^) =3.6 pulses/h=0.001pulses/p

the pulse rate for fiill scale = lOOOOOOms
the pulse width(Max.) = lOOOOOOms x 40% = 400000ms
but, the pulse width is 500ms Max.

* the pulse width(Max.) = SOOms

Case Range

Counting rate (pulse rate)

Setting pulse width

:3600m^/h (ImVs)
: l m ^
:0ms (automatically set)

the pulse rate

: 3600(m^/h) / l(m^) =3600 pulses/h=lpulses/p

the pulse rate for full scale = 1000ms
the pulse width(Max.) = 1000ms x 40% = 400ms
but, the pulse width that automatically set is 100ms Max.

* the pulse width (Max.) = 100ms

-