9. Calibration

9.1 Calibration Items

You can conduct the following in the calibration mode:

• Checks or calibrates the zero and span of the converter by using a simulation signal.

• Checks of the excitation current.

To change the mode to the calibration mode, see 0, “Mode Change.”

To check or change the zero and span of the converter and the excitation current value, follow the
procedure described below.

TOSHIBA_____________________________

However, these are already checked and calibrated when shipped from the factory.

Do not change these settings unless it is necessary to calibrate in the field.

Section

Configuration item

Display example

9.2.1

0 % flow rate calibration

FLOW CAL 0

9.2.2

50 % flow rate calibration

FLOW CAL 50

9.2.3

100 % flow rate calibration

FLOW CALI 00

9.2.4

Checking the excitation current value

EX CURR DSP

111

-