Defibrillator analyzer specifications, Energy output measurement – Fluke Biomedical 6000D Impulse User Manual

Page 45

Advertising
background image

Defibrillator/Transcutaneous Pacemaker Analyzer

Specifications

31

Defibrillator Analyzer Specifications

Energy Output Measurement

Compatible Defibrillator Waveshapes

.................Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic

Note

AC Pulsed Bi-Phasic waveform has not been approved in the United States.

Autoranged Measurement

....................................0.1 to 600 J

Accuracy

0.1 to 360 J .....................................................

±

(1 % of reading + 0.1 J)

360 to 600 J ....................................................

±

(1 % of reading + 0.1 J), typical

Note

For Pulsed Bi-Phasic defibrillator, specified accuracy is

±

(1.5 % of reading + 0.3 J) on both ranges.

Load resistance

Resistance...........................................................50

Ω

Accuracy..........................................................

±

1 %, non-inductive (<2

μ

H)

Pulse trigger level

.................................................20 V

Pulse width

Range..................................................................1.0 to 50.0 ms
Accuracy..............................................................

±

0.1 ms

Voltage

Range..................................................................20 to 5000 V
Accuracy..............................................................

±

(1 % of reading + 2 V)

Current

Range..................................................................0.4 to 100.0 A
Accuracy..............................................................

±

(1 % of reading + 0.1 A)