Defibrillator/Transcutaneous Pacemaker Analyzer

Creating Test Templates

57

5.

To define how Ansur calculates the limit values for
certain measurements, click the Operand field to
open a drop-down menu, as shown in Figure 2-35.

The operand can be set to any of the following:

Y – an absolute value

X + Y – an offset where the limit is calculated

as preset value + specified limit

X + (X * Y%) – calculated as a percentage

deviance from the preset value

When the operand is not an absolute limit, the

(dynamic) icon appears in the left column, as
shown in Figure 2-35. This icon indicates that the
limit will be calculated when the test is run.

fcz11.bmp

Figure 35. Changing the Operand in Expected Results