Arrhythmia wave test – Fluke Biomedical 6000D Impulse User Manual

Page 86

Advertising
background image

Impulse 6000D, 7000DP
Users Manual

72

Arrhythmia Wave Test

This test is typically used to verify the shock advisory
capability of a defibrillator in response to various
arrhythmia waveforms. Table 20 lists test custom
parameters.

Note

This is a visual / audible test; the Analyzer takes
no measurements during this test.

Table 20. Arrhythmia Wave Advisory Test Custom Parameters

Parameter Description

Arrhythmia

Selects the category of arrhythmia waveform to use for the test

Arrhythmia Type

Selects the type of arrhythmia to simulate

Poly VTach Type

Provides a choice of 5 types of Poly VTach waveforms when this type of arrhythmia is selected

Mono VTach Rate

BPM value is required when Mono VTach Rate is selected.

ECG Amplitude

The voltage amplitude to use for the arrhythmia simulation

Duration

How long the simulation should last. Setting can range from 1 to 60 seconds. With checkbox
Indefinite selected, duration is ignored, and the test must be manually stopped by the field user
(Stop in TEST GUIDE toolbar).

Atrial Width, Amplitude,
and Polarity

These atrial parameters are needed for some TV Paced arrhythmia simulations.

Ventricular Width,
Amplitude, and Polarity

These ventricular parameters are needed for some TV Paced arrhythmia simulations.