Synchronization time test – Fluke Biomedical 6000D Impulse User Manual

Page 77

Advertising
background image

Defibrillator/Transcutaneous Pacemaker Analyzer

Using Defibrillator Test Elements

63

Table 8. Charge Time Test Custom Parameters

Parameter Description

Preset Energy

The energy (in Joules) used for the test

The field user is prompted to set the defibrillator to this value.

ECG Waveform

The ECG waveform that the Analyzer should simulate during the test

Synchronization Time Test

The Synchronization Time test determines the ability of
the defibrillator to synchronize the discharge of its output
pulse with the simulated ECG waveform generated by the
Analyzer. Synchronization Time test measurements are

listed in Table 9. The Synchronization Time test custom
parameters are shown in Table 10.

Note

This test element supports the Auto Advance
Test Guide Setting.

Table 9. Synchronization Time Test Measurements

Measurement

Unit of Measure

Description

Sync Time

seconds

The delay between the top of the ECG wave and the discharging of the
defibrillator pulse

Energy

Joules

Amount of energy discharged by the defibrillator

Peak Voltage

Volts

Peak voltage detected during the discharge

Peak Current

Amperes

Peak current detected during the discharge