Users Manual

74

ECG R-Wave Test

The ECG R-Wave test checks the ability of the ECG
monitor to detect an R-Wave over a range of R-Wave
widths and amplitudes. Table 22 lists the custom
parameters used by the ECG R-Wave test.

Note

This is a visual / audible test; the Analyzer takes
no measurements during this test.

Table 22. ECG R-Wave Test Custom Parameters

Parameter Description

ECG Amplitude

Indicates the amplitude that the waveform should have

Performance Wave

Selects the type of Waveform to use for the test

Choices are:

Square: 0.125 or 2 Hz

Pulse: 30 or 60 BPM

Sine: 0.05, 0.5, 10, 40, 50, 60, 100, 150, 200 Hz

Triangle: 2 or 2.5 Hz

Duration

How long the simulation should last

Setting can range from 1 to 60 seconds. If checkbox Indefinite is selected, duration
is ignored, and the user must stop the test manually (Stop in TEST GUIDE toolbar).