Pacer demand mode test – Fluke Biomedical 6000D Impulse User Manual

Page 84

Advertising
background image

Impulse 6000D, 7000DP
Users Manual

70

Pacer Demand Mode Test

The Pacer Demand Mode test is used to verify demand
mode pacing over a range of BPM rates. This test is
conducted as a pass/fail test. Table 18 lists the custom
parameters used for the Pacer Demand Mode test.

Note

This is a visual / audible test; the Analyzer takes
no measurements during this test.

Table 18. Pacer Demand Mode Test Custom Parameters

Parameter Description

Input Jacks

Specifies where the pacer leads are attached on the Impulse 7000DP

Choose between pacer jacks or defib jacks.

Pacer Load

Defines load used for the test. If defib jacks are used for the test, a 50

Ω

load is used.

Brand

The brand of defibrillator/pacer being tested can be specified to optimize the accuracy of the test

DUT Rate

Expected pacer rate; field user prompted to set pacemaker to this rate

DUT Amplitude

Expected pacer amplitude; field user prompted to set the pacemaker output current to this value

ECG Simulation

This group of parameters establishes a range of BPM values that the test cycles through. The Starting
Rate sets the BPM for the start of the test. The Duration Per Step indicates how long the starting rate is
used. At the end of the duration time, the BPM is adjusted by the amount specified in the Auto
Increment / Decrement Amount parameter. The test concludes when the Ending Rate has been
reached.

Advertising
This manual is related to the following products: