Using pacemaker test elements(impulse 7000dp only), Pacer parameter test – Fluke Biomedical 6000D Impulse User Manual

Page 78

Advertising
background image

Impulse 6000D, 7000DP
Users Manual

64

Table Synchronization Time Test Measurements (cont)

Measurement

Unit of Measure

Description

Pulse Width 50%

milliseconds

The width of the pulse at 50% of its peak

Pulse Width 10%

milliseconds

The width of the pulse at 10% of its peak

Table 10. Synchronization Time Test Custom Parameters

Parameter Description

ECG Waveform

The ECG waveform the Analyzer should simulate during the test

Preset Energy

The energy (in Joules) used for the test

The field user is prompted to set the defibrillator to this value.

Using Pacemaker Test Elements
(Impulse 7000DP only)

Pacemaker tests confirm the basic operation of external
transcutaneous pacemakers by measuring various
pacemaker outputs and timing. These tests do not
operate with the Impulse 6000D analyzer.

Pacer Parameter Test

The Pacer Parameter test provides data about the
accuracy of a pacer’s output. Tables
11 and 12 list the Pacer Parameter test measurements
and custom parameters.