Defibrillator/Transcutaneous Pacemaker Analyzer

Specifications

35

Noise Immunity

Wave ................................................................... Sine
Line Frequency....................................................50 or 60 Hz (± 0.5 Hz)
Amplitude ............................................................0.0 to 10.0 (by 0.5) mV
Accuracy..............................................................± 5%

Transvenous Pacer Pulse Simulation

Widths

Range..............................................................0.1, 0.2, 0.5, 1.0, and 2.0 ms
Accuracy..........................................................±5 % of setting

Amplitudes...........................................................0 (off) and ±2, ±4, ±6, ±8, ±10, ±12, ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, and

±700 mV

Accuracy..............................................................± (10 % of setting + 0.2 mV)

Arrhythmia Selections

Pacer Interactive (Transcutaneous pacer, Impulse 7000DP only)

Demand...........................................................30 to 360 (by 1) BPM
Asynchronous
Non-Capture
Non-Function
Threshold (Interactive pacing simulation only) 10 to 250 (by 10) mA

Supraventricular

Atrial Fibrillation Coarse
Atrial Fibrillation fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia