Pacer sensitivity test – Fluke Biomedical 6000D Impulse User Manual

Page 81

Advertising
background image

Defibrillator/Transcutaneous Pacemaker Analyzer

Using Pacemaker Test Elements (Impulse 7000DP only)

67

Table 14. Pacer Refractory Test Custom Parameters

Parameter Description

Input Jacks

Specifies where the pacer leads are attached on the Impulse 7000

Choose between pacer jacks or defib jacks.

Pacer Load

Defines load used for the test

If defib jacks are used for the test, a 50

Ω

load is used.

Brand

The brand of defibrillator/pacer being tested can be specified to optimize the accuracy of the test.

DUT Rate

Expected pacer rate; field user prompted to set pacemaker to this rate

DUT Amplitude

Expected pacer amplitude; field user prompted to set the pacemaker output current to this value

Pacer Sensitivity Test

The Pacer Sensitivity test determines the threshold of
ECG amplitude required to trigger the pacemaker. Tables

15 and 16 list the Pacer Sensitivity test measurements
and custom parameters.

Table 15. Pacer Sensitivity Test Measurements

Measurement

Unit of Measure

Description

Sensitivity Threshold Amplitude

millivolts

The ECG amplitude that triggers the pacer