Defibrillator/Transcutaneous Pacemaker Analyzer

Specifications

33

Automated Defibrillator Test ECG Waves

Normal Sinus.......................................................30 to 300 (by 1) BPM
Ventricular Fibrillation..........................................Coarse and fine
Monomorphic Ventricular Tachycardia................120 to 300 (by 5) BPM
Polymorphic Ventricular Tachycardia..................5 types
Asystole...............................................................Flat line

ECG Waves

ECG General

Lead configuration...............................................12-lead simulation. RA, LL, LA, RL, V1-6 with independent outputs
Lead to lead impedance ......................................1000 Ω (nominal)
Rate accuracy .....................................................

±

1 % of nominal

ECG Amplitudes

Reference lead ....................................................Lead 1
Settings ...............................................................0.05 to 0.45 (by 0.05) mV

0.5 to 5.0 (by 0.5) mV

Accuracy..............................................................

±

2 % of setting, lead I and 2 Hz square wave

For performance waves and R-wave detection, other leads are proportional to Lead I in percentage per:

Lead I ..............................................................100
Lead II .............................................................150
Lead III ............................................................50
Leads V1 through V6 ......................................100

For normal sinus waves, other leads are proportional to Lead I in percentage per:

Lead I ..............................................................100
Lead II .............................................................150
Lead III ............................................................50