Transcutaneous Pacemaker Analyzer Specifications (Impulse 7000DP only)

Test Load Selections

Defibrillator Input

Fixed Load ..........................................................50 Ω
Accuracy .............................................................

±

1 %, non-inductive (<2

μ

H)

Power Rating.......................................................10 defib pulses of 360 J every 5 minutes

Pacemaker Input

Variable Load ......................................................50 to 1500

Ω

in 50

Ω

steps

Accuracy .............................................................

±

1 %, non-inductive (<2

μ

H)

Power Rating.......................................................5 W (average), 40 W (peak) @ 1000 Ω

Measurements

Manufacturer Specific Algorithms (plus general purpose default algorithm selection)

GE Responder (1500 & 1700)
MDE 300 (Medical Data Electronics)
Medtronic ERS/Physio Control LIFEPAK
MRL (Medical Research Laboratory/Welch Allyn)
Philips/Agilent/HP
Schiller Medical
ZOLL Medical
(plus a general purpose default algorithm selection)

Current

Range..................................................................4.00 to 250 mA
Accuracy .............................................................±(1% of reading + 0.02 mA)