Charge Time Test

The Charge Time test measures how long a defibrillator
takes to charge up to a specified energy level. Typically,
this test uses as a parameter the maximum energy level
available to the defibrillator. Table 7 lists the

measurements taken for this test. Custom parameters
available for the Charge Time tests are listed in Table 8.

Note

This test element supports the Auto Advance
Test Guide Setting.

Table 7. Charge Time Test Measurements

Measurement

Unit of Measure

Description

Charge Time

seconds

How long it took for the defibrillator to charge to the preset energy specified.

Energy

Joules

Amount of energy discharged by the defibrillator

Peak Voltage

Volts

Peak voltage detected during the discharge

Peak Current

Amperes

Peak current detected during the discharge

Pulse Width 50%

milliseconds

The width of the pulse at 50% of its peak

Pulse Width 10%

milliseconds

The width of the pulse at 10% of its peak