Performance wave simulation – Fluke Biomedical 6000D Impulse User Manual

Page 87

Advertising
background image

Defibrillator/Transcutaneous Pacemaker Analyzer

Using ECG Waveform Simulation Test Elements

73

Performance Wave Simulation

The Performance Wave simulation can be used to test
the integrity of a defibrillator monitor with a variety of
additional waveform shapes, such as square, triangle,
sine, and pulse. Table 21 lists the custom parameters for
this simulation.

Note

This is a visual / audible test; the Analyzer takes
no measurements during this test.

Table 21. Performance Wave Simulation Test Custom Parameters

Parameter Description

ECG Amplitude

Indicates the amplitude that the waveform should have

Performance Wave

Selects the type of Waveform to use for the test

Choices are:

Square: 0.125 or 2 Hz

Pulse: 30 or 60 BPM

Sine: 0.05, 0.5, 10, 40, 50, 60, 100, 150, 200 Hz

Triangle: 2 or 2.5 Hz

Duration

How long the simulation should last

Setting can range from 1 to 60 seconds. If checkbox Indefinite is selected, duration
is ignored, and the user must stop the test manually (Stop in TEST GUIDE toolbar).