Table 16. Pacer Sensitivity Test Custom Parameters

Parameter Description

Input Jacks

Specifies where the pacer leads are attached on the Impulse 7000DP

Choose between pacer jacks or defibrillator jacks.

Pacer Load

Defines load used for the test

If defibrillator jacks are used for the test, a fixed 50

Ω

load is used.

Brand

The brand of defibrillator/pacer being tested can be specified to optimize the accuracy of the test.

DUT Rate

Expected pacer rate; field user prompted to set pacemaker to this rate

DUT Amplitude

Expected pacer amplitude; field user prompted to set the pacemaker output current to this value

Waveform Type

The type of ECG waveform output to generate during the test

Waveform Width

The width of each waveform pulse

Waveform Polarity

The ECG waveform polarity that can be specified

ECG Pacer Interactive Test

The ECG Pacer Interactive test simulates a patient
response to a pacemaker. Table 17 lists the ECG Pacer
Interactive custom test parameters.

Note

This is a visual / audible test; the Analyzer takes
no measurements during this test.