Demand and asynchronous mode test – Fluke Biomedical 6000D Impulse User Manual

Page 53

Advertising
background image

Defibrillator/Transcutaneous Pacemaker Analyzer

Specifications

39

Pulse Rate

Range..................................................................5.0 to 800 PPM
Accuracy..............................................................±(0.5% of reading + 0.1 PPM)

Pulse Width

Range..................................................................1.00 to 100.0 ms
Accuracy..............................................................±(0.5% of reading + 0.01 ms)

Energy

Range..................................................................1 µJ to 2.00 J
Accuracy..............................................................±(4% of reading + 10 µJ)

Demand and Asynchronous Mode Test

Input Pacer pulse rates

.........................................30 to 200 PPM

ECG NSR wave

Rate.....................................................................10 to 300 (by 1) BPM
Amplitude ............................................................1 mV
Underdrive rate....................................................10 BPM minimum
Overdrive rate......................................................300 BPM maximum