Fluke Biomedical 6000D Impulse User Manual

Page 75

Advertising
background image

Defibrillator/Transcutaneous Pacemaker Analyzer

Using Defibrillator Test Elements

61

Table 5. Energy Measurement Test Measurements

Measurement

Unit of Measure

Description

Energy

Joules

Amount of energy discharged by the defibrillator

Peak Voltage

Volts

Peak voltage detected during the discharge

Peak Current

Amperes

Peak current detected during the discharge

Pulse Width 50%

milliseconds

The width of the pulse at 50% of its peak

Pulse Width 10%

milliseconds

The width of the pulse at 10% of its peak

Table 6. Energy Measurement Test Custom Parameters

Parameter Description

Preset Energy

The energy (in Joules) used for the test

The field user is prompted to set the defibrillator to this value.

ECG Waveform

The ECG waveform that the Analyzer should simulate during the test