Defib pulse repetition test – Fluke Biomedical 6000D Impulse User Manual

Page 65

Advertising
background image

Defibrillator/Transcutaneous Pacemaker Analyzer

Battery Performance Tests

51

5.

When charging is complete, discharge the
defibrillator. The first test result briefly appears in the
right pane of the TEST GUIDE window. A new
countdown timer message appears in the Test
results pane, showing the time remaining before the
next charge/discharge cycle.

6.

Wait for the countdown to conclude.

7.

Repeat steps 4 through 6 until the test is completed.

8.

To abort the text, click Stop in the TEST GUIDE
toolbar.

9.

When the test is fully complete, the measurement
results display in the Test results pane.

Defib Pulse Repetition Test

The Defib Pulse Repetition test is used to determine if a
battery-powered defibrillator can deliver a specified
number of discharges within a specified amount of time.

To run a Defib Pulse Repetition test:

1.

Connect the defibrillator to the Analyzer.

2.

Set the defibrillator to the energy level indicated in
the information block in the right pane of the TEST
GUIDE window.

3.

Click Start in the TEST GUIDE toolbar to start the
test. A progress window displays “Please wait…”
while Ansur configures the Analyzer.

4.

When configuration is complete, a prompt displays to
repeatedly charge and discharge the defibrillator,
along with a countdown timer showing the amount of
time available to complete the test..

5.

Discharge the defibrillator. Results briefly appear in
the right pane of the TEST GUIDE window.

6.

Continue to charge and discharge the defibrillator for
the remaining specified cycles or until the countdown
timer reaches zero.

7.

Click Stop in the TEST GUIDE toolbar to abort the
test.

8.

When the test is complete, the measurement results
display in the Test results pane.