Sensitivity test, Refractory period tests – Fluke Biomedical 6000D Impulse User Manual

Page 54

Advertising
background image

Impulse 6000D, 7000DP
Users Manual

40

Sensitivity Test

Automatic Interactive Threshold Detection

Compatible pacer rates .......................................30 to 120 PPM

ECG R wave:

Waveforms ..........................................................Square, Triangle, Sine
Width ...................................................................1 to 19 (by 1) ms

20 to 95 (by 5) ms
100 to 300 (by 25) ms

Accuracy .............................................................± 5% of setting
Amplitude ............................................................0.05 to 0.95 (by 0.05) mV

1.0 to 5.0 (by 0.5) mV

Accuracy .............................................................± 5% of setting

Refractory Period Tests

Paced Refractory Period

......................................20 to 500 ms

Sensed Refractory Period

....................................15 to 500 ms

Accuracy

................................................................±1 ms

Pacer pulse rate

....................................................20 to 200 PPM

ECG

Waveform............................................................Triangle wave
Pulse width..........................................................40 ms
Amplitude ............................................................1.0 mV