Asynchronous mode test, Using ecg waveform simulation testelements, Normal sinus wave simulation – Fluke Biomedical 6000D Impulse User Manual

Page 85: Using ecg waveform simulation test elements

Advertising
background image

Defibrillator/Transcutaneous Pacemaker Analyzer

Using ECG Waveform Simulation Test Elements

71

Asynchronous Mode Test

The Asynchronous Mode Test element requires the same
parameters as the Pacer Demand Mode Test, but this
test element is used for testing a non-demand mode
pacer. Refer to table 18 for the list of custom parameters
that this test uses.

Using ECG Waveform Simulation Test
Elements

Normal Sinus Wave Simulation

The Analyzer can generate a normal sinus wave between
10 and 360 beats per minute for output to a defibrillator
monitor. Table 19 lists the custom parameters for this
simulation.

Note

This is a visual / audible test; the Analyzer takes
no measurements during this test.

Table 19. Normal Sinus Wave Simulation Test Custom Parameters

Parameter Description

ECG Amplitude

The voltage amplitude to use during the simulation of the normal sinus wave

Normal Sinus Waveform BPM

The BPM rate to use for the test

Duration

How long the simulation should last

Setting can range from 1 to 60 seconds. If the “Indefinite” checkbox is checked, the
duration is ignored, and the test must be manually stopped by the field user (Stop in
TEST GUIDE toolbar).