Defibrillator/Transcutaneous Pacemaker Analyzer

Using Pacemaker Test Elements (Impulse 7000DP only)

65

Table 11. Pacer Parameter Test Measurements

Measurement

Unit of Measure

Description

Pacer Rate

Pulses/minute

Number of pacer pulses detected by the Analyzer during the test

Pulse Amplitude

milliamperes

Peak current detected during the test

Pulse Width

milliseconds

Pacer pulse width as measured by the Analyzer

Pacer Energy

millijoules

Pacemaker energy output

Table 12. Pacer Parameter Test Custom Parameters

Parameter Description

Input Jacks

Specifies where the pacer leads are attached on the Impulse 7000DP

Choose between pacer jacks or defib jacks.

Pacer Load

Defines load used for the test

If defib jacks are used for the test, a 50

Ω

load is used.

Brand

The brand of defibrillator/pacer being tested can be specified to optimize the accuracy of the test.

DUT Rate

Expected pacer rate; field user prompted to set pacemaker to this rate

DUT Amplitude

Expected pacer amplitude; field user prompted to set the pacemaker output current to this value