Defib pulse repetition test – Fluke Biomedical 6000D Impulse User Manual

Page 91

Advertising
background image

Defibrillator/Transcutaneous Pacemaker Analyzer

Using Battery Performance Test Elements

77

Defib Pulse Repetition Test

The Defib Pulse Repetition test is used to determine if a
battery-powered defibrillator can deliver a specified

number of discharges within a specified amount of time.
Tables 26 and 27 list the Defib Pulse Repetition test
measurements and custom parameters.

Table 26. Defib Pulse Repetition Test Measurements

Measurement

Unit of Measure

Description

Energy

Joules

Amount of energy discharged by the defibrillator

The test stores multiple energy measurements – one per discharge.

Table 27. Defib Pulse Repetition Test Custom Parameters

Parameter Description

Preset Energy

The energy (in Joules) used for the test

The field user is prompted to set the defibrillator to this value.

Number of Pulses

The number of discharges to be completed during the test

Number of Minutes

The maximum time allowed for this test to be completed