TECHNICAL DATA, MOUNTING AND ELECTRICAL CONNECTIONS

23

23


Calculated outputs

1. Analog (0-20, 4-20 mA) outputs:

Any 2 of the 7 calculated massecuite parameters (per pan).

2. Digital (ON / OFF) outputs:

1. DO1: Warning on approaching seeding (programmable).

2. DO2: Opening the seeding valve for a selected time interval.


NOTE:
All calculated and measured data can be also transmitted via digital communication.

Process interface

Inputs

1. Analog current

8 channels

Symmetrical, galvanically isolated

Range

0(4)-20 mA, keyboard selectable

Input impedance

100 Ohm

Common mode voltage

50 V max.

2. RTD (Pt 100) (option)

2 channels

4 wire connecting mode

3. Digital (ON / OFF)

8 channels

Signal sources:

passive, contact or open collector

active, + 24 V

dc

automatic signal detection

Outputs

1. Analog current

4 channels

Galvanically isolated

Range

0(4)-20 mA, keyboard selectable

Max. load

600 Ohm

2. Digital ON / OFF)

4 channels

Isolated open collector,

overvoltage and short circuit protected

3. Power supply for transmitters

1 X 24 Vdc, 200 mA max.

4. Device Status (ON/OFF)

Operating / Error


Communication

1. Serial (COM1, COM2, COM3)

3 ports

Galvanically isolated

Standards

RS232, RS422, RS485

(keyboard selectable)

Control signals

CTS, RTS

(keyboard selectable)

Baud rates

1200…38400 Baud

(keyboard selectable)

Length of cable

RS232

15 m max.

RS422 /485

1200 m max.

Protocols

K-PATENTS refractometer protocol

MODBUS SLAVE (ASCII, RTU)


2. Ethernet 10 6 100 Base T

Connector

RJ45

Protocols

TCP / IP, MODBUS TCP,

UDP / IP (K-PATENTS refractometer protocol)

Front panel


Display 5,7 ‘ QVGA 320x240 graphic LCD

Keyboard

Membrane switches with flexible foil cover

LED indicators

Power, Run, Alarm

Power supply

100…240 V ac, 50 / 60 Hz; 10-25 VA

24 (18…30) Vdc


Advertising